Recall Z-2172-2018— BioMerieux SA
Class II · Terminated
Class II FDA medical device recall by BioMerieux SA, initiated 2018-03-23, terminated 2020-06-23. FDA's recall record names no specific 510(k) clearance for this event. Product: Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria. This enzyme is characteristic of the genus Neisseria and most species of Pseudomonas. It enables the differentiation of Gram (-) bacilli (1, 2). A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.. Reason: A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers..
- Recalling firm
- BioMerieux SA
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-03-23
- Terminated
- 2020-06-23
- Firm location
- Marcy L'Etoile, N/A, France
Product description
Oxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria. This enzyme is characteristic of the genus Neisseria and most species of Pseudomonas. It enables the differentiation of Gram (-) bacilli (1, 2). A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
Reason for recall
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Data sourced from openFDA. This site is unofficial and independent of the FDA.