Recall Z-2171-2018— BioMerieux SA

Class II · Terminated

Class II FDA medical device recall by BioMerieux SA, initiated 2018-03-23, terminated 2020-06-23. FDA's recall record names no specific 510(k) clearance for this event. Product: Color Gram 2 (COLOR GRAM 2 - F), these stains are used to stain bacterial and fungal microorganisms for direct examination, according to the Gram method. Dye and chemical solution stains for medical purposes are mixtures of synthetic or natural dyes or nondye chemicals in solutions used in staining cells and tissues for diagnostic histopathology, cytopathology, or hematology.. Reason: A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers..

Recalling firm
BioMerieux SA
Classification
Class II
Status
Terminated
Initiated
2018-03-23
Terminated
2020-06-23
Firm location
Marcy L'Etoile, N/A, France

Product description

Color Gram 2 (COLOR GRAM 2 - F), these stains are used to stain bacterial and fungal microorganisms for direct examination, according to the Gram method. Dye and chemical solution stains for medical purposes are mixtures of synthetic or natural dyes or nondye chemicals in solutions used in staining cells and tissues for diagnostic histopathology, cytopathology, or hematology.

Reason for recall

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Data sourced from openFDA. This site is unofficial and independent of the FDA.