Recall Z-2170-2026— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2026-03-31. It names 3 FDA 510(k) clearances: K200917, K181830, K172822. Product: Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).. Reason: Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2026-03-31
Firm location
Best, Netherlands

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).

Reason for recall

Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.