Recall Z-2170-2018— BioMerieux SA

Class II · Terminated

Class II FDA medical device recall by BioMerieux SA, initiated 2018-03-23, terminated 2020-06-23. FDA's recall record names no specific 510(k) clearance for this event. Product: The PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI¿ Color Gram instrument sprays the PREVI¿ Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI¿ Color Gram platform that is also indicated on the reagent label. An automated slide stainer is a device used to stain histology, cytology, and hematology slides for diagnosis.. Reason: A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers..

Recalling firm
BioMerieux SA
Classification
Class II
Status
Terminated
Initiated
2018-03-23
Terminated
2020-06-23
Firm location
Marcy L'Etoile, N/A, France

Product description

The PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI¿ Color Gram instrument sprays the PREVI¿ Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI¿ Color Gram platform that is also indicated on the reagent label. An automated slide stainer is a device used to stain histology, cytology, and hematology slides for diagnosis.

Reason for recall

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Data sourced from openFDA. This site is unofficial and independent of the FDA.