Recall Z-2166-2024— Horiba Instruments Incorporated

Class II · Ongoing

Class II FDA medical device recall by Horiba Instruments Incorporated, initiated 2024-03-28. FDA's recall record names no specific 510(k) clearance for this event. Product: HORIBA custom configured fluorescence instrument, modular Fluorolog-QM. Reason: HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure..

Recalling firm
Horiba Instruments Incorporated
Classification
Class II
Status
Ongoing
Initiated
2024-03-28
Firm location
Piscataway, NJ, United States

Product description

HORIBA custom configured fluorescence instrument, modular Fluorolog-QM

Reason for recall

HORIBA custom configured fluorescence instrument has a laser safety defect. This custom configuration of the modular Instrument removed the primary light source and the excitation monochromator. A plastic cap was installed to plug the optical port, where a more durable beam stop should have been used. The defect may allow the class 4 laser beam to burn (or transmit) through the plastic cap to exit the protective enclosure.

Data sourced from openFDA. This site is unofficial and independent of the FDA.