Recall Z-2165-2026— Dexcom, Inc.

Class II · Ongoing

Class II FDA medical device recall by Dexcom, Inc., initiated 2026-04-14. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System. Reason: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information..

Recalling firm
Dexcom, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-04-14
Firm location
San Diego, CA, United States

Product description

Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System

Reason for recall

A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.

Data sourced from openFDA. This site is unofficial and independent of the FDA.