Recall Z-2165-2018— BioMerieux SA

Class II · Terminated

Class II FDA medical device recall by BioMerieux SA, initiated 2018-03-23, terminated 2020-06-23. FDA's recall record names no specific 510(k) clearance for this event. Product: The VITEK 2 Neisseria-Haemophilus identification card (NH) is intended for use with the VITEK 2 Systems for the automated identification of most clinically significant fastidious organisms. The VITEK 2 NH identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.. Reason: A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers..

Recalling firm
BioMerieux SA
Classification
Class II
Status
Terminated
Initiated
2018-03-23
Terminated
2020-06-23
Firm location
Marcy L'Etoile, N/A, France

Product description

The VITEK 2 Neisseria-Haemophilus identification card (NH) is intended for use with the VITEK 2 Systems for the automated identification of most clinically significant fastidious organisms. The VITEK 2 NH identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.

Reason for recall

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Data sourced from openFDA. This site is unofficial and independent of the FDA.