Recall Z-2163-2026— Dexcom, Inc.

Class II · Ongoing

Class II FDA medical device recall by Dexcom, Inc., initiated 2026-04-14. It names 2 FDA 510(k) clearances: K243214, K240902. Product: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System. Reason: A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information..

Recalling firm
Dexcom, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-04-14
Firm location
San Diego, CA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System

Reason for recall

A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.