Recall Z-2159-2019— DRG Instruments GmbH

Class II · Terminated

Class II FDA medical device recall by DRG Instruments GmbH, initiated 2019-05-21, terminated 2020-05-29. It names one FDA 510(k) clearance: K052649. Product: DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013 Product Usage: The Salivary Testosterone ELISA (SlV-3013) is an enzymatic in vitro Assay for the detection of Testosterone in human Saliva. Testosterone levels are measured in fertility studies in men and women. Fertility problems are not diagnosed on the outcome of Testosterone alone but accompanied by other diagnostic means. Reason: Salivary assay may observe a higher percentage of samples with 0 pg/ml testosterone or results which are too low. Saliva samples in the low measuring range of the kit(< 50 pg/ml) were assayed significantly lower.

Recalling firm
DRG Instruments GmbH
Classification
Class II
Status
Terminated
Initiated
2019-05-21
Terminated
2020-05-29
Firm location
Marburg, N/A, Germany

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

DRG Salivary Testosterone ELISA- IVD for the detection of Testerone in human saliva Catalog # SLV-3013 Product Usage: The Salivary Testosterone ELISA (SlV-3013) is an enzymatic in vitro Assay for the detection of Testosterone in human Saliva. Testosterone levels are measured in fertility studies in men and women. Fertility problems are not diagnosed on the outcome of Testosterone alone but accompanied by other diagnostic means

Reason for recall

Salivary assay may observe a higher percentage of samples with 0 pg/ml testosterone or results which are too low. Saliva samples in the low measuring range of the kit(< 50 pg/ml) were assayed significantly lower