Recall Z-2158-2019— Polymer Technology Systems, Inc.
Class III · Terminated
Class III FDA medical device recall by Polymer Technology Systems, Inc., initiated 2019-06-07, terminated 2020-07-08. It names one FDA 510(k) clearance: K151545. Product: Henry Schein Panels Lipid Panel test strip: IVD Lipid Panel Test Strips Ref Number: 570-0414 Product Usage: PTS Panels Lipid Panel test strips measure total cholesterol, HDL (high density lipoprotein) cholesterol, and triglycerides in whole blood with the CardioChek PA or the CardioChek Plus professional analyzers, and provide a quantitative result. A TC/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA and CardioChek Plus analyzers. A MEMo Chip¿ is provided with each package of test strips and must be properly inserted into the analyzer before any test can be run. The MEMo Chip contains the test name, calibration curve, lot number and test strip expiration date. After the test strip is inserted into the analyzer and blood applied to the test strip, test results are displayed in as little as 90 seconds.. Reason: The MEMo chips provided with lot P901 may be mislabeled, the MEMo Chip may display Q901 on the analyzer LCD screen, instead of P901.
- Recalling firm
- Polymer Technology Systems, Inc.
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2019-06-07
- Terminated
- 2020-07-08
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Henry Schein Panels Lipid Panel test strip: IVD Lipid Panel Test Strips Ref Number: 570-0414 Product Usage: PTS Panels Lipid Panel test strips measure total cholesterol, HDL (high density lipoprotein) cholesterol, and triglycerides in whole blood with the CardioChek PA or the CardioChek Plus professional analyzers, and provide a quantitative result. A TC/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol are calculated by the CardioChek PA and CardioChek Plus analyzers. A MEMo Chip¿ is provided with each package of test strips and must be properly inserted into the analyzer before any test can be run. The MEMo Chip contains the test name, calibration curve, lot number and test strip expiration date. After the test strip is inserted into the analyzer and blood applied to the test strip, test results are displayed in as little as 90 seconds.
Reason for recall
The MEMo chips provided with lot P901 may be mislabeled, the MEMo Chip may display Q901 on the analyzer LCD screen, instead of P901