Recall Z-2157-2025— Bridge to Life Ltd

Class II · Ongoing

Class II FDA medical device recall by Bridge to Life Ltd, initiated 2025-06-12. It names one FDA 510(k) clearance: K191006. Product: Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250. Reason: The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time..

Recalling firm
Bridge to Life Ltd
Classification
Class II
Status
Ongoing
Initiated
2025-06-12
Firm location
Northbrook, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP), Part Number REF BTLE-1250

Reason for recall

The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.