Recall Z-2156-2019— SenTec AG
Class II · Terminated
Class II FDA medical device recall by SenTec AG, initiated 2019-05-02, terminated 2020-12-16. It names 3 FDA 510(k) clearances: K151329, K101690, K041548. Product: SenTec Membrane Changer Insert [5 pcs - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.]. Reason: There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure..
- Recalling firm
- SenTec AG
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-05-02
- Terminated
- 2020-12-16
- Firm location
- Therwil, N/A, Switzerland
510(k) clearances named in this recall
3 on FDA's record- K151329 — SenTec Digital Monitor, OxiVenT Sensor, V-Sign Sensor, Staysite Adhesive Pad, V-STATS PC Software including V-CareNet
- K101690 — SENTEC DIGITAL MONITOR, V-SIGN SENSOR AND V-SIGN SENSOR 2 WITH DIGITAL SENSOR ADAPER CABLE
- K041548 — SENTEC DIGITAL MONITOR MODEL #SDM
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
SenTec Membrane Changer Insert [5 pcs - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.]
Reason for recall
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
Data sourced from openFDA. This site is unofficial and independent of the FDA.