Recall Z-2155-2012— Stryker Spine
Class II · Terminated
Class II FDA medical device recall by Stryker Spine, initiated 2012-04-23, terminated 2013-11-13. FDA's recall record names no specific 510(k) clearance for this event. Product: LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-FRANCE; Distributed in the USA by: Stryker Spine. 2 Pearl Ct. Allendale, NJ 07401. Provides access to the thoracic and lumbar spine from a posterior approach via a small incision.. Reason: Stryker received three reports involving blades of the Luxor Retractor becoming disassembled from the base causing the retractor to collapse insitu..
- Recalling firm
- Stryker Spine
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-04-23
- Terminated
- 2013-11-13
- Firm location
- Allendale, NJ, United States
Product description
LUXOR RETRACTOR BASE; Catalog Number: 48250030; Manufactured by: STRYKER Spine SAS; Z.I MARTICOT-33610-CESTAS-FRANCE; Distributed in the USA by: Stryker Spine. 2 Pearl Ct. Allendale, NJ 07401. Provides access to the thoracic and lumbar spine from a posterior approach via a small incision.
Reason for recall
Stryker received three reports involving blades of the Luxor Retractor becoming disassembled from the base causing the retractor to collapse insitu.