Recall Z-2151-2024— Baxter Healthcare Corporation
Class II · Ongoing
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2024-05-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Centrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries.Multiple model numbers of Centrella Max mattresses and pro+ mattresses. (1) Centrella Max Surface, REF P7922A01; (2) Centrella Max Surface, REF P7922A02; (3) Centrella Max Surface with X-Ray, REF P7922A03; (4) Centrella Max,Narrow, Rental, REF P7922ARENT01; (5) Centrella Pro+ 36" Surface with X-Ray, REF P7923A01; (6) Centrella Pro+ 40" Surface with X-Ray, REF P7923A02; (7) Centrella Pro+36" Surface, REF P7923A03; (8) Centrella Pro+ 40" Surface, REF P7923A04; (9) Pro+ MRS Surface for Hillrom 900 with X-Ray, REF P006800A01; (10) 80" MRS with X-Ray AU/NZ REF P006800A03; (11) Pro+ MRS Surface for Versacare with X-Ray, REF P3255A01; (12) Pro+ MRS Surface for Versacare, REF P3255A02; (13) Pro + MRS Surface for Versacare with X-Ray, REF P3255ARENT01; (14) Pro+ 36"MRS Surface with X-Ray, REF P7924A01; (15) Pro+ 40" MRS Surface with X-Ray, REF P7924A02; (16) Pro+ 36" MRS Surface, REF P7924A03; (17) Pro+ 40" MRS Surface, REF P7924A04; (18) Pro+ 36" MRS Rental Surface with X-Ray, REF P7924ARENT01.. Reason: There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-05-08
- Firm location
- Deerfield, IL, United States
Product description
Centrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries.Multiple model numbers of Centrella Max mattresses and pro+ mattresses. (1) Centrella Max Surface, REF P7922A01; (2) Centrella Max Surface, REF P7922A02; (3) Centrella Max Surface with X-Ray, REF P7922A03; (4) Centrella Max,Narrow, Rental, REF P7922ARENT01; (5) Centrella Pro+ 36" Surface with X-Ray, REF P7923A01; (6) Centrella Pro+ 40" Surface with X-Ray, REF P7923A02; (7) Centrella Pro+36" Surface, REF P7923A03; (8) Centrella Pro+ 40" Surface, REF P7923A04; (9) Pro+ MRS Surface for Hillrom 900 with X-Ray, REF P006800A01; (10) 80" MRS with X-Ray AU/NZ REF P006800A03; (11) Pro+ MRS Surface for Versacare with X-Ray, REF P3255A01; (12) Pro+ MRS Surface for Versacare, REF P3255A02; (13) Pro + MRS Surface for Versacare with X-Ray, REF P3255ARENT01; (14) Pro+ 36"MRS Surface with X-Ray, REF P7924A01; (15) Pro+ 40" MRS Surface with X-Ray, REF P7924A02; (16) Pro+ 36" MRS Surface, REF P7924A03; (17) Pro+ 40" MRS Surface, REF P7924A04; (18) Pro+ 36" MRS Rental Surface with X-Ray, REF P7924ARENT01.
Reason for recall
There is a potential for the top cover of the hospital bed mattress to delaminate as a result of improper cleaning and lack of adequate details on the appropriate use of cleaning agents containing bleach.
Data sourced from openFDA. This site is unofficial and independent of the FDA.