Recall Z-2151-2023— Datascope Corp.

Class I · Ongoing

Class I FDA medical device recall by Datascope Corp., initiated 2023-06-05. It names one FDA 510(k) clearance: K181122. Product: Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution. Reason: An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy..

Recalling firm
Datascope Corp.
Classification
Class I
Status
Ongoing
Initiated
2023-06-05
Firm location
Mahwah, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution

Reason for recall

An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.