Recall Z-2146-2012— Encore Medical, Lp

Class II · Terminated

Class II FDA medical device recall by Encore Medical, Lp, initiated 2012-07-16, terminated 2012-12-04. It names one FDA 510(k) clearance: K931665. Product: djo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell or fracture repair. Reason: One lot of the 010-55-030, 6.5 mm Cancellous Bone Screw was recalled because a product complaint identified a mislabeling of the device. Although the package label indicates Size 30mm, the screw dimension is actually 25mm..

Recalling firm
Encore Medical, Lp
Classification
Class II
Status
Terminated
Initiated
2012-07-16
Terminated
2012-12-04
Firm location
Austin, TX, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

djo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell or fracture repair

Reason for recall

One lot of the 010-55-030, 6.5 mm Cancellous Bone Screw was recalled because a product complaint identified a mislabeling of the device. Although the package label indicates Size 30mm, the screw dimension is actually 25mm.