Recall Z-2140-2016— Hospira Inc.

Class II · Terminated

Class II FDA medical device recall by Hospira Inc., initiated 2013-03-08, terminated 2016-11-07. It names one FDA 510(k) clearance: K043256. Product: The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.. Reason: Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion..

Recalling firm
Hospira Inc.
Classification
Class II
Status
Terminated
Initiated
2013-03-08
Terminated
2016-11-07
Firm location
Lake Forest, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.

Reason for recall

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

Data sourced from openFDA. This site is unofficial and independent of the FDA.