Recall Z-2140-2014— Ortho-Clinical Diagnostics
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2014-07-07, terminated 2018-05-21. It names 2 FDA 510(k) clearances: K063144, K031924. Product: VITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.) The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).. Reason: Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-07-07
- Terminated
- 2018-05-21
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
2 on FDA's record- K063144 — VITROS 5,1 FS CHEMISTRY SYSTEM WITH ENGEN LABORATORY AUTOMATION SYSTEM
- K031924 — VITROS 5,1 FS CHEMISTRY SYSTEM, VITROS CHEMISTRY PRODUCTS DHDL REAGENT PACK, CALIBRATOR KIT 19, FS CALIBRATOR 1
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.) The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).
Reason for recall
Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.