Recall Z-2139-2020— Radiometer Medical ApS
Class II · Terminated
Class II FDA medical device recall by Radiometer Medical ApS, initiated 2019-11-21, terminated 2024-11-08. It names one FDA 510(k) clearance: K092686. Product: ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.. Reason: The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors..
- Recalling firm
- Radiometer Medical ApS
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-11-21
- Terminated
- 2024-11-08
- Firm location
- Bronshoj, N/A, Denmark
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
Reason for recall
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.