Recall Z-2136-2020— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2020-01-29, terminated 2021-04-28. It names one FDA 510(k) clearance: K042291. Product: ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.. Reason: Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2020-01-29
Terminated
2021-04-28
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.

Reason for recall

Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.