Recall Z-2135-2020— MOTEK MEDICAL B.V.
Class II · Ongoing
Class II FDA medical device recall by MOTEK MEDICAL B.V., initiated 2019-07-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Body Weight Support (BWS) Light; Model No. RB011; a component of C-Mill, Grail, M-Gait - Product Usage: is designed to facilitate functional gait training of subjects with gait and/or balance impairment by providing dynamic partial vertical unloading during walking on a treadmill. The device will be used as an assistive device for gait deficiency rehabilitation and research purposes.. Reason: The firm discovered a potential manufacturing deviation of the yoke of the Body Weight Support (BWS) Light that in rare occasions may result in the yoke detaching from the cable guide..
- Recalling firm
- MOTEK MEDICAL B.V.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2019-07-29
- Firm location
- Amsterdam, N/A, Netherlands
Product description
Body Weight Support (BWS) Light; Model No. RB011; a component of C-Mill, Grail, M-Gait - Product Usage: is designed to facilitate functional gait training of subjects with gait and/or balance impairment by providing dynamic partial vertical unloading during walking on a treadmill. The device will be used as an assistive device for gait deficiency rehabilitation and research purposes.
Reason for recall
The firm discovered a potential manufacturing deviation of the yoke of the Body Weight Support (BWS) Light that in rare occasions may result in the yoke detaching from the cable guide.