Recall Z-2131-2025— GE Medical Systems, LLC
Class II · Ongoing
Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2025-05-23. FDA's recall record names no specific 510(k) clearance for this event. Product: SIGNA Architect AIR. Reason: GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual..
- Recalling firm
- GE Medical Systems, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-05-23
- Firm location
- Waukesha, WI, United States
Product description
SIGNA Architect AIR
Reason for recall
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Data sourced from openFDA. This site is unofficial and independent of the FDA.