Recall Z-2123-2025— Ortho-Clinical Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2025-06-10. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);. Reason: Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias)..

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-06-10
Firm location
Rochester, NY, United States

Product description

VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);

Reason for recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Data sourced from openFDA. This site is unofficial and independent of the FDA.