Recall Z-2122-2016— Covidien, LLC

Class II · Terminated

Class II FDA medical device recall by Covidien, LLC, initiated 2016-06-06, terminated 2017-10-18. It names one FDA 510(k) clearance: K152794. Product: Covidien Cytosponge Cell Collection Kit. Contains Cytosponge Cell Collection Device, Specimen Jar with preservative. Item code CYTO-KIT-R Gastroenterology: Cytosponge Cell Collection Device is indicated for use in the collection and retrieval of surface cells in the esophagus. Currently the device is only available for clinical studies.. Reason: Medtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal string during withdrawal from the patient's esophagus..

Recalling firm
Covidien, LLC
Classification
Class II
Status
Terminated
Initiated
2016-06-06
Terminated
2017-10-18
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Covidien Cytosponge Cell Collection Kit. Contains Cytosponge Cell Collection Device, Specimen Jar with preservative. Item code CYTO-KIT-R Gastroenterology: Cytosponge Cell Collection Device is indicated for use in the collection and retrieval of surface cells in the esophagus. Currently the device is only available for clinical studies.

Reason for recall

Medtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal string during withdrawal from the patient's esophagus.

Data sourced from openFDA. This site is unofficial and independent of the FDA.