Recall Z-2119-2016— Ortho-Clinical Diagnostics
Class III · Terminated
Class III FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2016-05-24, terminated 2018-06-08. It names one FDA 510(k) clearance: K052867. Product: VITROS Chemistry Products Calibrator Kit 29, REF/Product Code 680 2344, used in conjunction with VITROS Chemistry Products dTIBC Reagent Generation 30 (GEN 30), REF 680 2001, and Assay Data Disk (ADD), REF/Product Code 6801876 (Unique Device No. 10758750001576), Data Release Versions (DRVs) 5870 through 5883 inclusive; IVD --- Ortho-Clinical Diagnostics For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 29 is used to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent.. Reason: Ortho Clinical Diagnostics started receiving customer complaints for biased results when using VITROS dTIBC Reagent GEN 30 product. Ortho's investigation confirmed that incorrect calibration mathematics were assigned to the Calibrator Kit 29 Lots 2915 and 2995 supporting dTIBC Reagent GEN 30 product..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2016-05-24
- Terminated
- 2018-06-08
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS Chemistry Products Calibrator Kit 29, REF/Product Code 680 2344, used in conjunction with VITROS Chemistry Products dTIBC Reagent Generation 30 (GEN 30), REF 680 2001, and Assay Data Disk (ADD), REF/Product Code 6801876 (Unique Device No. 10758750001576), Data Release Versions (DRVs) 5870 through 5883 inclusive; IVD --- Ortho-Clinical Diagnostics For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 29 is used to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of total iron-binding capacity (TIBC) using VITROS Chemistry Products dTIBC Reagent.
Reason for recall
Ortho Clinical Diagnostics started receiving customer complaints for biased results when using VITROS dTIBC Reagent GEN 30 product. Ortho's investigation confirmed that incorrect calibration mathematics were assigned to the Calibrator Kit 29 Lots 2915 and 2995 supporting dTIBC Reagent GEN 30 product.
Data sourced from openFDA. This site is unofficial and independent of the FDA.