Recall Z-2118-2020— Clinical Innovations, LLC

Class II · Terminated

Class II FDA medical device recall by Clinical Innovations, LLC, initiated 2019-05-30, terminated 2020-08-24. It names one FDA 510(k) clearance: K131781. Product: Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICAL INNOVATIONS Manufacturer: Clinical Innovations, LLC* * *Reorder Number: UM750* * *LOT 181183* * *." Each device is accompanied by Instruction for Use which is similarly labeled.. Reason: An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury..

Recalling firm
Clinical Innovations, LLC
Classification
Class II
Status
Terminated
Initiated
2019-05-30
Terminated
2020-08-24
Firm location
Murray, UT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICAL INNOVATIONS Manufacturer: Clinical Innovations, LLC* * *Reorder Number: UM750* * *LOT 181183* * *." Each device is accompanied by Instruction for Use which is similarly labeled.

Reason for recall

An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury.