Recall Z-2116-2023— Abiomed, Inc.
Class I · Ongoing
Class I FDA medical device recall by Abiomed, Inc., initiated 2023-06-14. FDA's recall record names no specific 510(k) clearance for this event. Product: Impella CP with SmartAssist intravascular micro axial blood pump, Product Numbers 0048-0024, 0048-0045, 1000080. Reason: There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility..
- Recalling firm
- Abiomed, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-06-14
- Firm location
- Danvers, MA, United States
Product description
Impella CP with SmartAssist intravascular micro axial blood pump, Product Numbers 0048-0024, 0048-0045, 1000080
Reason for recall
There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.
Data sourced from openFDA. This site is unofficial and independent of the FDA.