Recall Z-2116-2014— Pentax Medical Company

Class II · Terminated

Class II FDA medical device recall by Pentax Medical Company, initiated 2014-06-12, terminated 2015-11-17. It names 2 FDA 510(k) clearances: K041396, K041397. Product: EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gastroscope FG-36UX Ultrasound Fiber Gastroscope Used to provide optical and sonographic visualization of and therapeutic access to the Upper Gastrointestinal Tract.. Reason: Aspiration needles used in combination with Ultrasound Gastroscopes can derail from the accessory elevator at the distal end of the endoscope. This may result in patient injury if the Instruction for Use is not followed carefully by the user..

Recalling firm
Pentax Medical Company
Classification
Class II
Status
Terminated
Initiated
2014-06-12
Terminated
2015-11-17
Firm location
Montvale, NJ, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

EG-3630U Ultrasound Video Gastroscope EG-3830UT Ultrasound Video Gastroscope EG-3870UTK Ultrasound Video Gastroscope FG-36UX Ultrasound Fiber Gastroscope Used to provide optical and sonographic visualization of and therapeutic access to the Upper Gastrointestinal Tract.

Reason for recall

Aspiration needles used in combination with Ultrasound Gastroscopes can derail from the accessory elevator at the distal end of the endoscope. This may result in patient injury if the Instruction for Use is not followed carefully by the user.