Recall Z-2115-2016— Philips Medical Systems (Cleveland) Inc

Class II · Terminated

Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2016-06-21, terminated 2018-07-02. It names one FDA 510(k) clearance: K033326. Product: Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body.. Reason: Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel..

Recalling firm
Philips Medical Systems (Cleveland) Inc
Classification
Class II
Status
Terminated
Initiated
2016-06-21
Terminated
2018-07-02
Firm location
Cleveland, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ingenuity Core, model # 728326; Intended to produce cross- sectional images of the body.

Reason for recall

Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.