Class I FDA medical device recall by Abiomed, Inc., initiated 2023-06-14. FDA's recall record names no specific 510(k) clearance for this event. Product: Impella 5.5 with SmartAssist intravascular micro axial blood pump, Product Numbers 0550-0008 and 1000100. Reason: There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility..
- Recalling firm
- Abiomed, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-06-14
- Firm location
- Danvers, MA, United States
Product description
Impella 5.5 with SmartAssist intravascular micro axial blood pump, Product Numbers 0550-0008 and 1000100
Reason for recall
There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.