Recall Z-2112-2023— Abiomed, Inc.

Class I · Ongoing

Class I FDA medical device recall by Abiomed, Inc., initiated 2023-06-14. FDA's recall record names no specific 510(k) clearance for this event. Product: Impella 5.0 intravascular micro axial blood pump, Product Number 005062. Reason: There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility..

Recalling firm
Abiomed, Inc.
Classification
Class I
Status
Ongoing
Initiated
2023-06-14
Firm location
Danvers, MA, United States

Product description

Impella 5.0 intravascular micro axial blood pump, Product Number 005062

Reason for recall

There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

Data sourced from openFDA. This site is unofficial and independent of the FDA.