Recall Z-2110-2016— Philips Medical Systems (Cleveland) Inc
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2016-06-14, terminated 2018-07-02. It names one FDA 510(k) clearance: K033357. Product: Big Bore Radiology tomography X-ray system scanner, model number 728244. The Big Bore Radiology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.. Reason: Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel..
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-06-14
- Terminated
- 2018-07-02
- Firm location
- Cleveland, OH, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Big Bore Radiology tomography X-ray system scanner, model number 728244. The Big Bore Radiology scanner is a whole body computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Reason for recall
Socket Head Cap Screws (SHCS) used in the Z-Axis Adjustment Plate were found to be made of stainless steel rather than the specified alloy steel.
Data sourced from openFDA. This site is unofficial and independent of the FDA.