Recall Z-2109-2017— Toshiba American Medical Systems Inc

Class II · Terminated

Class II FDA medical device recall by Toshiba American Medical Systems Inc, initiated 2017-02-01, terminated 2018-06-06. It names one FDA 510(k) clearance: K152785. Product: Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F. Reason: It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on..

Recalling firm
Toshiba American Medical Systems Inc
Classification
Class II
Status
Terminated
Initiated
2017-02-01
Terminated
2018-06-06
Firm location
Tustin, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F

Reason for recall

It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.

Data sourced from openFDA. This site is unofficial and independent of the FDA.