Recall Z-2107-2012— Smiths Medical ASD, Inc.
Class I · Terminated
Class I FDA medical device recall by Smiths Medical ASD, Inc., initiated 2012-02-14, terminated 2012-11-14. It names one FDA 510(k) clearance: K111386. Product: Smiths Medical, Medfusion¿ Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78. Designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates.. Reason: Smiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe alarm. When the device exhibits this failure mode, visual and audible alarms will sound and the device ceases operation..
- Recalling firm
- Smiths Medical ASD, Inc.
- Classification
- Class I
- Status
- Terminated
- Initiated
- 2012-02-14
- Terminated
- 2012-11-14
- Firm location
- Saint Paul, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Smiths Medical, Medfusion¿ Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78. Designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates.
Reason for recall
Smiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe alarm. When the device exhibits this failure mode, visual and audible alarms will sound and the device ceases operation.
Data sourced from openFDA. This site is unofficial and independent of the FDA.