Recall Z-2107-2012— Smiths Medical ASD, Inc.

Class I · Terminated

Class I FDA medical device recall by Smiths Medical ASD, Inc., initiated 2012-02-14, terminated 2012-11-14. It names one FDA 510(k) clearance: K111386. Product: Smiths Medical, Medfusion¿ Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78. Designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates.. Reason: Smiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe alarm. When the device exhibits this failure mode, visual and audible alarms will sound and the device ceases operation..

Recalling firm
Smiths Medical ASD, Inc.
Classification
Class I
Status
Terminated
Initiated
2012-02-14
Terminated
2012-11-14
Firm location
Saint Paul, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Smiths Medical, Medfusion¿ Syringe Infusion Pump, Model 4000, Software Version V1.0 & V1.1, Reorder Numbers 4000-0100-50, 4000-0100-95, 4000-0101-50, 4000-0101-51, and 4000-0101-78. Designed for use in care areas for the administration of fluids requiring precisely controlled infusion rates.

Reason for recall

Smiths Medical has identified a software anomaly in the Medfusion 4000 Syringe Infusion Pump that causes a device history log corruption and triggers a Watchdog Fail-Safe alarm. When the device exhibits this failure mode, visual and audible alarms will sound and the device ceases operation.