Recall Z-2106-2019— Comet Technologies Usa Inc

Class II · Terminated

Class II FDA medical device recall by Comet Technologies Usa Inc, initiated 2018-10-01, terminated 2024-06-13. FDA's recall record names no specific 510(k) clearance for this event. Product: YXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control. Reason: Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door..

Recalling firm
Comet Technologies Usa Inc
Classification
Class II
Status
Terminated
Initiated
2018-10-01
Terminated
2024-06-13
Firm location
Shelton, CT, United States

Product description

YXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control

Reason for recall

Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.

Data sourced from openFDA. This site is unofficial and independent of the FDA.