Recall Z-2105-2015— ArthroCare Corporation
Class II · Terminated
Class II FDA medical device recall by ArthroCare Corporation, initiated 2015-06-02, terminated 2016-06-27. It names 2 FDA 510(k) clearances: K142999, K070374. Product: EVac 70 XTRA with Integrated Cable and PROcise EZ View with Integrated Cable Product Usage: PLASMA WAND devices are indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: adenoidectomy, cysts, head, neck, oral, and sinus surgery, mastoidectomy, myringotomy with effective hemorrhage control, nasal airway obstruction be reduction of hypertrophic nasal turbinates, nasopharyngeal/laryngeal indications including tracheal procedures, laryngeal polypectomy, and laryngeal lesion debulking, neck mass, papilloma keloids, submucosal palatal shrinkage, submucosal tissue shrinkage, tonsillectomy (including palatine tonsils), traditional uvulopalatoplasty (RAUP), tumors, and tissue in the uvula/soft palate for the treatment of snoring. It is intended for procedures using a conductive media, such as normal saline or Ringers lactate.. Reason: Wands inspected with the adaptor fixture were released to the field not meeting the dielectric (Hipot) specifications..
- Recalling firm
- ArthroCare Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-06-02
- Terminated
- 2016-06-27
- Firm location
- Austin, TX, United States
510(k) clearances named in this recall
2 on FDA's record- K142999 — EVac 70 Xtra Plasma Wand with Integrated Cable, PROcise XP Plasma Wand with Integrated Cable
- K070374 — ARTHROCARE ENT PLASMA WAND
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EVac 70 XTRA with Integrated Cable and PROcise EZ View with Integrated Cable Product Usage: PLASMA WAND devices are indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: adenoidectomy, cysts, head, neck, oral, and sinus surgery, mastoidectomy, myringotomy with effective hemorrhage control, nasal airway obstruction be reduction of hypertrophic nasal turbinates, nasopharyngeal/laryngeal indications including tracheal procedures, laryngeal polypectomy, and laryngeal lesion debulking, neck mass, papilloma keloids, submucosal palatal shrinkage, submucosal tissue shrinkage, tonsillectomy (including palatine tonsils), traditional uvulopalatoplasty (RAUP), tumors, and tissue in the uvula/soft palate for the treatment of snoring. It is intended for procedures using a conductive media, such as normal saline or Ringers lactate.
Reason for recall
Wands inspected with the adaptor fixture were released to the field not meeting the dielectric (Hipot) specifications.