Recall Z-2104-2014— Terumo Cardiovascular Systems Corporation
Class II · Terminated
Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2014-05-23, terminated 2015-02-24. It names one FDA 510(k) clearance: K080034. Product: Terumo Three-stage Venous Return Cannulae 28 Fr with 3/8" flare, 30 Fr with 3/8" connector and 30 Fr with 3/8" flare. Product Usage: Venous Return Cannula is indicated for venous drainage during cardiopulmonary bypass surgery for dual cannulation of the superior and inferior vena cava.. Reason: During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications.
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-05-23
- Terminated
- 2015-02-24
- Firm location
- Ann Arbor, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Terumo Three-stage Venous Return Cannulae 28 Fr with 3/8" flare, 30 Fr with 3/8" connector and 30 Fr with 3/8" flare. Product Usage: Venous Return Cannula is indicated for venous drainage during cardiopulmonary bypass surgery for dual cannulation of the superior and inferior vena cava.
Reason for recall
During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications
Data sourced from openFDA. This site is unofficial and independent of the FDA.