Recall Z-2099-2021— Randox Laboratories Ltd.

Class II · Terminated

Class II FDA medical device recall by Randox Laboratories Ltd., initiated 2021-05-10, terminated 2021-11-10. FDA's recall record names no specific 510(k) clearance for this event. Product: NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115. Reason: Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples.

Recalling firm
Randox Laboratories Ltd.
Classification
Class II
Status
Terminated
Initiated
2021-05-10
Terminated
2021-11-10
Firm location
Crumlin (North), N/A, Ireland

Product description

NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

Reason for recall

Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples