Recall Z-2098-2025— GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Class II · Ongoing
Class II FDA medical device recall by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, initiated 2025-06-20. It names one FDA 510(k) clearance: K953801. Product: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System. Reason: Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury..
- Recalling firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-06-20
- Firm location
- Tirat Carmel, N/A, Israel
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
Reason for recall
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.