Recall Z-2097-2015— Maquet Cardiovascular Us Sales, Llc

Class II · Terminated

Class II FDA medical device recall by Maquet Cardiovascular Us Sales, Llc, initiated 2015-06-10, terminated 2016-04-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.. Reason: There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision..

Recalling firm
Maquet Cardiovascular Us Sales, Llc
Classification
Class II
Status
Terminated
Initiated
2015-06-10
Terminated
2016-04-18
Firm location
Wayne, NJ, United States

Product description

Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.

Reason for recall

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.