Recall Z-2094-2023— CareFusion 303, Inc.
Class II · Ongoing
Class II FDA medical device recall by CareFusion 303, Inc., initiated 2023-05-22. FDA's recall record names no specific 510(k) clearance for this event. Product: BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517. Reason: Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock..
- Recalling firm
- CareFusion 303, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-05-22
- Firm location
- San Diego, CA, United States
Product description
BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517
Reason for recall
Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock.
Data sourced from openFDA. This site is unofficial and independent of the FDA.