Recall Z-2082-2017— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2017-04-03, terminated 2018-05-31. FDA's recall record names no specific 510(k) clearance for this event. Product: HDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.. Reason: Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2017-04-03
Terminated
2018-05-31
Firm location
Brea, CA, United States

Product description

HDL Cholesterol Reagent (HDL 2 x 200), Catalog No. A15625. Intended for quantitative determination of HDL cholesterol in the high density lipoprotein fraction of human serum or plasma.

Reason for recall

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.