Recall Z-2081-2016— Alcon Research, Ltd.

Class II · Terminated

Class II FDA medical device recall by Alcon Research, Ltd., initiated 2016-05-24, terminated 2017-04-13. It names one FDA 510(k) clearance: K121555. Product: CENTURION Vision System. Reason: The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed..

Recalling firm
Alcon Research, Ltd.
Classification
Class II
Status
Terminated
Initiated
2016-05-24
Terminated
2017-04-13
Firm location
Fort Worth, TX, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CENTURION Vision System

Reason for recall

The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.

Data sourced from openFDA. This site is unofficial and independent of the FDA.