Recall Z-2080-2017— Beckman Coulter Inc.

Class II · Terminated

Class II FDA medical device recall by Beckman Coulter Inc., initiated 2017-04-03, terminated 2018-05-31. It names one FDA 510(k) clearance: K093458. Product: Enzymatic Creatinine Reagent (CR-E 2 x 200), Catalog No. A60298. Intended for the quantitative determination of creatinine (CR-E) concentration in human serum, plasma or urine.. Reason: Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results..

Recalling firm
Beckman Coulter Inc.
Classification
Class II
Status
Terminated
Initiated
2017-04-03
Terminated
2018-05-31
Firm location
Brea, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Enzymatic Creatinine Reagent (CR-E 2 x 200), Catalog No. A60298. Intended for the quantitative determination of creatinine (CR-E) concentration in human serum, plasma or urine.

Reason for recall

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.