Recall Z-2079-2025— Becton, Dickinson and Company, BD Bio Sciences
Class II · Ongoing
Class II FDA medical device recall by Becton, Dickinson and Company, BD Bio Sciences, initiated 2025-06-03. It names one FDA 510(k) clearance: K040026. Product: REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.. Reason: Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples..
- Recalling firm
- Becton, Dickinson and Company, BD Bio Sciences
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-06-03
- Firm location
- Milpitas, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
REF: 666289 , BD FACS, 7-Color Setup Beads, 25 Runs, CE, IVD. Used in flow cytometry for fluorescence compensation.
Reason for recall
Due to customer complaints, errors in spectral overlap factors table for 7-color setup beads may affect accurate setup of the flow cytometry instrument prior to analyzing samples.