Recall Z-2079-2012— Getinge USA Inc

Class II · Terminated

Class II FDA medical device recall by Getinge USA Inc, initiated 2012-05-25, terminated 2016-05-18. FDA's recall record names no specific 510(k) clearance for this event. Product: Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128. Reason: Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error..

Recalling firm
Getinge USA Inc
Classification
Class II
Status
Terminated
Initiated
2012-05-25
Terminated
2016-05-18
Firm location
Rochester, NY, United States

Product description

Getinge brand --- large capacity, floor loading jet spray, washers for mechanical washing, intermediate level, thermal disinfection and drying of moisture and temperature stable Hospital case carts, sterilization containers, material handling carts, stands and utensils. --- Model numbers Getinge 9120, 9122, 9125, and 9128

Reason for recall

Getinge Disinfection AB is voluntarily correcting Washer Disinfectors, Getinge Electrically Heated 9100-Series manufactured between 2009-05-04 through 09-26-2011. This action is based on a report from the ground fault breaker manufacturer of a production error.