Recall Z-2078-2014— Dental EZ Group Star Dental Division
Class II · Terminated
Class II FDA medical device recall by Dental EZ Group Star Dental Division, initiated 2014-06-13, terminated 2015-02-24. It names one FDA 510(k) clearance: K960260. Product: Titan T 5K Motor Product Usage: Low-speed motor is used with various attachments to perform a variety of dental procedures including caries removal, finishing, polishing, pin setting and prophy.The motor has a speed range of 100 to 5000 rpm capability.. Reason: The housing of 5K motor was laser marked incorrectly as a 20K motor. The recall was expanded in September 2014 to include additional lot #2251181..
- Recalling firm
- Dental EZ Group Star Dental Division
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-06-13
- Terminated
- 2015-02-24
- Firm location
- Lancaster, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Titan T 5K Motor Product Usage: Low-speed motor is used with various attachments to perform a variety of dental procedures including caries removal, finishing, polishing, pin setting and prophy.The motor has a speed range of 100 to 5000 rpm capability.
Reason for recall
The housing of 5K motor was laser marked incorrectly as a 20K motor. The recall was expanded in September 2014 to include additional lot #2251181.
Data sourced from openFDA. This site is unofficial and independent of the FDA.