Recall Z-2067-2020— Merit Medical Systems, Inc.

Class II · Terminated

Class II FDA medical device recall by Merit Medical Systems, Inc., initiated 2020-02-13, terminated 2021-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 1402 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).. Reason: Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the procedure to be rescheduled..

Recalling firm
Merit Medical Systems, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-02-13
Terminated
2021-02-26
Firm location
South Jordan, UT, United States

Product description

DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 1402 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

Reason for recall

Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the procedure to be rescheduled.

Data sourced from openFDA. This site is unofficial and independent of the FDA.