Recall Z-2065-2026— Physio-Control, Inc.

Class II · Ongoing

Class II FDA medical device recall by Physio-Control, Inc., initiated 2026-04-13. FDA's recall record names no specific 510(k) clearance for this event. Product: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122. Reason: Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy..

Recalling firm
Physio-Control, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-04-13
Firm location
Redmond, WA, United States

Product description

LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122

Reason for recall

Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

Data sourced from openFDA. This site is unofficial and independent of the FDA.