Recall Z-2064-2026— Erbe Medical, LLC
Class II · Ongoing
Class II FDA medical device recall by Erbe Medical, LLC, initiated 2026-03-31. It names one FDA 510(k) clearance: K190469. Product: ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222. Reason: Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure..
- Recalling firm
- Erbe Medical, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-03-31
- Firm location
- Tempe, AZ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
Reason for recall
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.